公司名称:XXX
电话:XXX
手机:XXX
联系人:XXX
地址:XXX
1. 每天对设备的运行状况进行不少于2次的巡检和记录。每天必须检测硬度、余氯、电导率,每月检测细菌、内毒素,水质合格后可投入使用。
Inspect and record the equipment in no less than two times a day. The hardness, chlorine and electrical conductivity must be tested every day, and the bacteria and the endotoxin are tested every month and can be put into use after the water quality is qualified.
2. 耗材更换要及时。PP棉滤芯每3个月更换一次,反渗透膜每2~3年更换一次,如产水量下降30%以上或产品水电导率高于10μ s/cm以上也需更换反渗透膜。
The consumables need to be replaced in time. PP cotton filter once every 3 months replacement, reverse osmosis membrane replaced once every 2 ~ 3 years, such as water rate fell by more than 30% or product water conductivity is higher than 10 mu s/cm above also need to change the reverse osmosis membrane.
3. 设备需定期消毒,在微生物检测超过干预值时需消毒。
The equipment needs to be disinfected regularly and disinfected when the detection of microorganism exceeds the intervention value.
4. 运行管理人员必须熟悉水处理工艺和设备的运行要求和技术指标。
The running manager must be familiar with the operation requirements and technical specifications of water treatment technology and equipment.
操作人员必须经过技术培训和生产实践,了解处理工艺,熟悉本岗位设施、设备的运行要求和技术指标考核合格后方可上岗。未经培训的非专业人员不可私自操作此设备。
Operators must pass through technical training and production practice, understand the treatment process, familiar with the operation requirements of this position facilities, equipment and technique index inspection qualified rear can mount guard. Untrained non-professional personnel shall not operate the equipment privately.
5. 设备故障,必须由专业操作人员联系厂家工程师排除,不可私自拆卸设备。
The equipment failure must be made by the professional operator to contact the manufacturer's engineer to remove the equipment.
6. 应定时检查电缆线路是否有损坏的迹象,保证有可靠的接地线,如有损坏必须立即更换。
Check to see if there are any signs of damage to the cable line, and ensure that there is a reliable grounding line, which must be replaced immediately if there is any damage.
7. 设备在消毒及运行时会有局部高温,非经培训的专业人员不可触碰设备。消毒冲洗未完成不得进行透析,有紧急透析患者,需中途停止消毒,必须完成冲洗程序,并对水质进行检测,水质合格后方可使用。
The equipment should be sterilized and run at a local temperature. Non-trained professionals must not touch the equipment. Disinfection washing unfinished no dialysis, dialysis patients of emergency, need to stop halfway to disinfection, washing procedures must be completed and and to test the water quality, water quality qualified rear can use.
在消毒时,科室必须在透析大厅及设备旁边悬挂明显警示牌,说明系统正在消毒中,谨防烫伤。
In the case of disinfection, the room must be hung by the dialysis hall and the equipment and the warning signs are displayed, indicating that the system is being disinfected, and is wary of scalding.